DDM 1000 Precision Thermometer for Laboratory and Quality Assurance
The DDM 1000 is designed for precise temperature measurements as demanded in calibration-laboratory use and quality assurance. The accuracy is adapted to the uncertainties of fix-point-cells for precise sensor calibrations. Pt 10-, Pt-25- and Pt 100-Sensors can be used
 at the same time. The resolution of up to 0,02 mK often is needed in physical and chemical laboratories.
Product Description
Specification
Order Information
Main Features
• Accuracy: 0,1 mK
• Resolution: 0,02 mK
• Range: -220 °C to +1200 °C
• Pt100, -25 and Pt10-Sensors
• User defined coefficients for ITS 90 and EN60751
• Fast measurements: up to 8/s
• Excellent long-time stability
• Compensation of self heating error
• No impairment of the measurement results through parasitic thermovoltage
• Extensible up to 81 measuring channels
• True 4-wire-technique with Lemo-plugs
• High EMC
• High data reliability in terms of sensor coefficients
Delivery:
supertronsensing162078.gif
Model : DDM 1000
 Order No                                   5020-1000             
supertronsensing162077.jpg
 Measuring range  -200°C up to 962°C (according to ITS-90)
 -200°C up to 850°C (according to DIN EN60751)
 Resolution  1 mK
 Measuring uncertainty  10 mK (optionally 5 mK in the range of -50AC up to 250AC)
 Measuring channels  2 (optionally up to 81 with external multiplexers)
 Sensors  Pt-100/Pt-25/Pt-10 in a four-conductor-technology
 Connectors  Lemo 1S, quadripolar, in front panel
 Measuring current  about 0,5 mA switched DC (about 1 mA for Pt-25)
 Measuring time  about 1 sec. per channel, about 0,1 sec. per channel with
 reduced resolution
 Long-term stability  5 mK / year
 Temperature coeff.  1 mK/K
 Display  LCD, LED backlight, 2 lines ¨¤ 16 characters, 9 mm height
 Units  °C, °F, K, Ohm
 Interface  RS-232, isolated
 Operation  menu orientated, alternatively controllable by your PC
 Memory  128 KB (optionally 512 KB)
 Suppy  230 V, 50 Hz / 115 V, 60 Hz about 15 VA
 Size  260 x 80 x 240 (width x height x depth in mm)
 Weight  about 2,5 kg
 Technical data  The following specifications are only valid for a nominal
operating voltage and a environmental temperature of 23AC.
 Available accessoires
 - external multiplexers with 8 or 16 channels
  (up to 5 switches may be cascade-connected)
- PC software for operation of instrument and measuring data acquisition
- temperature sensors
supertronsensing162076.jpg supertronsensing162073.gif supertronsensing162072.gif supertronsensing162071.gif supertronsensing162070.gif supertronsensing162069.gif supertronsensing162068.gif supertronsensing162067.gif supertronsensing162066.gif supertronsensing162065.gif supertronsensing162064.gif supertronsensing162063.gif supertronsensing162062.gif supertronsensing162061.gif supertronsensing162060.gif supertronsensing162059.gif supertronsensing162058.gif supertronsensing162057.gif supertronsensing162056.gif supertronsensing162055.gif supertronsensing162054.gif supertronsensing162053.gif supertronsensing162052.gif supertronsensing162051.gif supertronsensing162050.gif supertronsensing162049.gif supertronsensing162048.gif supertronsensing162047.gif supertronsensing162046.gif supertronsensing162045.gif supertronsensing162044.gif supertronsensing162043.gif supertronsensing162042.gif supertronsensing162041.gif supertronsensing162037.gif supertronsensing162036.gif supertronsensing162035.gif supertronsensing162033.gif supertronsensing162031.gif supertronsensing162028.gif supertronsensing162026.gif supertronsensing162025.gif supertronsensing162023.gif supertronsensing162021.gif supertronsensing162019.gif supertronsensing162017.gif supertronsensing162015.gif supertronsensing162013.gif supertronsensing162011.gif supertronsensing162009.gif supertronsensing162006.gif supertronsensing162004.gif supertronsensing162003.gif supertronsensing162001.gif