HygroGen Series
Set-point stabilisation time:
Reference probe :
Desiccant:
Chamber volume:
Housing / Dimensions :
Ambient conditions :
Weight :
Power supply:
Certification marks :
2 minutes (35 / 80 % rh humidity change, 23 °C)
HygroClip S1 calibrated at 5, 23 & 50 °C,
10, 35, 65 & 95 % rh,verifiable to SCS standard (UK: UKAS)
Molecular sieve, user refillable
Approx. 2 litres
Stainless steel / 455 x 420 x 212 mm (max.)
15 kg
110/230 VAC, 3 A 50/60 Hz
EN61326: 1998, EN61000-3-2: 2000,
EN61000-3-3: 1995, EN61010-1: 2001
HygroGen 0
Maximum control range :
Minimum control range :
5…95 % rh
10…90 % rh
in a range of 10…30 °C
Control stability :
< ± 0.2 % rh
Temperature gradient :
< ~0.2 °C (5...50°C) <0.1at 23 °C
Specification
Precision of reference probe:
External interfaces :
< ±1.0 % rh (10…95 % rh) ±0.2 K
Ethernet, ROTRONIC DIO
(two additional connections fitted)
HygroGen 1A
This self-contained humidity generator for calibration of humidity measuringdevices only needs a mains power socket and is light enough at 17 kg to be portable for on-site use.
The wide control range of 5…95 % rh / 5…50 °C and extremely high control stability make the HygroGen indispensable wherever there is a need for fast and precise calibration.
HygroGen 2A
The HygroGen 2A is equipped with a sampling loop and pump for connection of a reference dew point mirror. The maximum dew point is limited by the ambient temperature to which the lines of the sampling loop are exposed.
HygroGen 0
The HygroGen 0 does not include active temperature control of the chamber. The ambient conditions therefore define the chamber temperature, with carefully designed ventilators ensuring the best-possible homogeneity.
Max. 80 % rh at 10…30 °C
Saturator :
Humidifier with front filling. Low water level warning on controller display
Set-point stabilisation time:
Reference probe :
Desiccant:
Chamber volume:
Housing / Dimensions :
Ambient conditions :
Weight :
Power supply:
Certification marks :
2 minutes (35 / 80 % rh humidity change, 23 °C),
10 minutes (23 / 45 °C temperature change)
HygroClip S1 calibrated at 5, 23 & 50 °C, 10, 35, 65 & 95 % rh,verifiable to SCS standard (UK: UKAS)
Molecular sieve, user refillable
Approx. 2 litres
Stainless steel / 455 x 420 x 212 mm (max.)
17 kg(hygrogen 1A) 17.5kg(hygrogen 2A)
110/230 VAC, 3 A 50/60 Hz
EN61326: 1998, EN61000-3-2: 2000,
EN61000-3-3: 1995, EN61010-1: 2001
HygroGen 1A Hyrowgen 2A
Maximum control range :
Minimum control range :
5…95 % rh and 5…50 °C (optional 0 °C ...60 °C)
10…90 % rh in a range of 10…50 °C
Control stability :
< ± 0.2 % rh, 0.1 °C (at 23 °C) 0.2 °C (complete range)
Temperature gradient :
< ~0.2 °C (5...50°C) <0.1at 23 °C
Precision of reference probe:
External interfaces :
< ±1.0 % rh (10…95 % rh) ±0.2 K
Ethernet, ROTRONIC DIO
(two additional connections fitted)
Max. 80 % rh for temperatures up to 31 °C,
decreasing linearly to 50% rh at 40 °C
Indoor use only.
Altitude to 2,000 m above sea level
Saturator :
Humidifier with front filling. Low water level warning on controller display
Delivery: