Moisture
Sensor
|
Range: -110°C to +20°C dewpoint ( -166°F to +68°F dewpoint )
Accuracy: ±2°C temperature corrected Response time: 95% of step change in 3 min.
Repeatability: 0.8°C
Sample flow: >1 LPM
|
| Temperature Sensor |
Range:
-40°C to +70°C Accuracy: ±2°C |
| Ergonomics | Easy and quick installation, no tools required. |
| Electrical |
5VDC
to 28VDC two wire loop operation 4mA to 20mA
Housing isolated from 4-20mA Loop Measurement resolution of moisture sensor: 1pF resulting in less than 0.01°C 5 pin M12 industrial connector IP68 when mating connector fully engaged Mating connectors with molded cables approved for Hazardous Location use available
Two conductor cable min. 24AWG (for total cable length >1mile min. 20AWG) To meet CE requirements cable must be shielded.
Built-in digital I/O for controlling relays
|
| Mechanical |
All
316 Stainless Steel and PTFE wetted parts for fast response time.
Enclosure all stainless steel construction IP67 protection when installed Dimensions (overall) including connector- L:
3.25" (82.6mm) Dia: 1.62" (41.3mm)
process threads 3/4"-16 and 14mm x 1.25mm Weight Total: 6 oz (0.17 Kg) Pressure: 5000 PSIA (345 Bar) |
| Temperature Range |
PLMa
Analyzer: -20°C to +60°C
Electronics: -40°C to +85°C
Moisture Sensor: -20°C to +60°C
|
| Miscellaneous Features |
NIST
traceable calibrations
Units of measure: °C & °F dewpoint, ppmV, ppmW, µB H2O vapor pressure,
grams of H2O / m3 and Lbs H2O /106 standard cubic feet in Natural Gas. (All units autoscaled)
Digital & Analog: Two Wire 4/20mA loop with HART® Transmitter stores all HART setup |