supertronsensing237079.jpg
High precision data logger for temperature (±0,03), relative humidity, dew point, air flow with memory for 6.000 measurements. The instrument is designed for application of the quality assurance, e.g. testing climatic chambers, calibration thermostats, oven...or thermoelectrical processes. For the battery-powered instrument we provide a separate power supply as an option. The measuring interval is adjustable between 1 second and 30 minutes. By using our precision Pt100-probes with sensor calibration code you will reach the on page 12 descripted system accuracy. The P655-LOG measuring system is the perfect standard to test subordinated instruments or systems. For traceability we recommend to order those instruments with a calibration certificate.
Product Description
Specification
Order Information
Main Features
• RS232 interface galvanically isolated
• Optional SmartGraph Windows Software for graphic and tabular documentation
• High measuring accuracy
• Integrated calibration function for simple compensation of sensor tolerances
• Physical 1-point, 2-point or 3-point calibration function
• Measuring channels are freely assignable
• Recording maximum, minimum, hold and average values
• Integrated sensor holder for one hand operation
• Mains operation possible
• Simultaneous display of two measured values
• Differential temperature display
Delivery:
supertronsensing237077.gif
 Order No                                   5000-0655L      
 Inputs  Pt100, thermocouple: type K, J, L, N, R, S, T, relative
 humidity, air flow
 Pt100  -200°C...+850°C
 Thermocouple  -200°C...+1760°C (EN 60584-1)
 Ohm  0...400 Ohm 
 Humidity  0%...100% rH
 Flow  0...40 m/s
 Pt100  ±0,03°C from -50°C...+199,99°C
 ±0,05°C from -200°C...+200°C
 otherwise 0,1% of reading
 Thermocouple R, S  ±1,0°C +0,1% of reading
 Thermocouple K, J, L, N, T  ±0,2°C from 0°C...+200°C
 ±0,5°C to 1000°C
 ±1,0°C remaining range
 Humidity  ±1,5% rH
 Flow  0,5% of reading
 Ohm  0,5% of reading
 Resolution  0,01°C from -200°C...+200°C
 remaining range 0,1°C, 0,1%, 0,01 m/s
 Memory  6000 measurements (3.000 per channel)
 Measuring interval  Adjustable: 1 second...30 minutes
 Connectors  8-pole DIN 45326
 Working temperature  0°C...+40°C
 Display  2-line LCD
 Housing  plastic (ABS)
 Dimensions  200 x 85 x 40 mm (LxWxH)
 Weight   300 g
 Power supply      9 V battery  
P655-Log
supertronsensing237076.jpg supertronsensing237073.gif supertronsensing237072.gif supertronsensing237071.gif supertronsensing237070.gif supertronsensing237069.gif supertronsensing237068.gif supertronsensing237067.gif supertronsensing237066.gif supertronsensing237065.gif supertronsensing237064.gif supertronsensing237063.gif supertronsensing237062.gif supertronsensing237061.gif supertronsensing237060.gif supertronsensing237059.gif supertronsensing237058.gif supertronsensing237057.gif supertronsensing237056.gif supertronsensing237055.gif supertronsensing237054.gif supertronsensing237053.gif supertronsensing237052.gif supertronsensing237051.gif supertronsensing237050.gif supertronsensing237049.gif supertronsensing237048.gif supertronsensing237047.gif supertronsensing237046.gif supertronsensing237045.gif supertronsensing237044.gif supertronsensing237043.gif supertronsensing237042.gif supertronsensing237041.gif supertronsensing237037.gif supertronsensing237036.gif supertronsensing237035.gif supertronsensing237033.gif supertronsensing237031.gif supertronsensing237028.gif supertronsensing237026.gif supertronsensing237025.gif supertronsensing237023.gif supertronsensing237021.gif supertronsensing237019.gif supertronsensing237017.gif supertronsensing237015.gif supertronsensing237013.gif supertronsensing237011.gif supertronsensing237009.gif supertronsensing237006.gif supertronsensing237004.gif supertronsensing237003.gif supertronsensing237001.gif